| Model Number |
24686 |
| Medical Device Problem Code |
Material Puncture/Hole (1504)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/01/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that balloon shaft hole occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, it was noted that the balloon had a small hole at the base of the shaft.The procedure was completed with another of the same device.No patient complications reported and the patient's status was fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr.: gladiator 10 x 40, 75cm was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 12mm proximal from the proximal markerband.The rated burst pressure for this device is 14 atmospheres.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
|
| |
|
Event or Problem Description
|
|
It was reported that balloon shaft hole occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, it was noted that the balloon had a small hole at the base of the shaft.The procedure was completed with another of same device.No patient complications reported and the patient's status was fine.
|
| |
|
Search Alerts/Recalls
|