• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Code Material Puncture/Hole (1504)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/01/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that balloon shaft hole occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, it was noted that the balloon had a small hole at the base of the shaft.The procedure was completed with another of the same device.No patient complications reported and the patient's status was fine.
 
Additional Manufacturer Narrative
Device evaluated by mfr.: gladiator 10 x 40, 75cm was returned for analysis.A visual examination identified that the balloon was not folded which indicates that the balloon was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 12mm proximal from the proximal markerband.The rated burst pressure for this device is 14 atmospheres.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination identified no kinks or damage to the shaft which may have potentially contributed to the complaint incident.A visual and tactile examination identified no damage to the tip which may have potentially contributed to the complaint incident.No other issues were identified during the product analysis.
 
Event or Problem Description
It was reported that balloon shaft hole occurred.A 10.0 x40, 75cm gladiator elite balloon catheter was selected for use.However, it was noted that the balloon had a small hole at the base of the shaft.The procedure was completed with another of same device.No patient complications reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10722203
Report Number2134265-2020-14163
Device Sequence Number3006371
Product Code LIT
UDI-Device Identifier08714729809517
UDI-Public08714729809517
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/15/2023
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0025516324
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/20/2020
Initial Date Received by Manufacturer 10/01/2020
Supplement Date Received by Manufacturer10/27/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date11/12/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-