• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD CARIBE LTD. NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE; ANESTHESIA CONDUCTION NEEDLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD CARIBE LTD. NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE; ANESTHESIA CONDUCTION NEEDLE Back to Search Results
Model Number 405180
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event or Problem Description
It was reported that 63 needles spinal s/su 25ga 3-1/2in quincke were found discolored before use.The following information was provided by the initial reporter: "reported issue: discoloration".
 
Additional Manufacturer Narrative
H6: investigation summary 63 samples were received by our quality team for evaluation.Upon visual inspection and evaluation of all 63 samples, it was observed that the samples were in product specification; therefore, the incident could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.Based on the investigation, the samples did not contain discoloration, variation in color, foreign matter, nor cloudiness condition and meet product specification requirements.Since the incident could not be verified a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event or Problem Description
It was reported that 63 needles spinal s/su 25ga 3-1/2in quincke were found discolored before use.The following information was provided by the initial reporter: "reported issue: discoloration".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEEDLE SPINAL S/SU 25GA 3-1/2IN QUINCKE
Common Device Name
ANESTHESIA CONDUCTION NEEDLE
Manufacturer (Section D)
BD CARIBE LTD.
road 31
k.m. 24.3
juncos
MDR Report Key10723559
Report Number2618282-2020-00088
Device Sequence Number1679636
Product Code BSP
UDI-Device Identifier30382904051809
UDI-Public30382904051809
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
PMA/510(K) Number
PREAMENDMENT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source other,user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Other
Device Expiration Date04/30/2025
Device Model Number405180
Device Catalogue Number405180
Device Lot Number0125473
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/09/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer09/29/2020
Initial Report FDA Received Date10/22/2020
Supplement Report FDA Received Date10/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-