| Model Number |
405180 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that 63 needles spinal s/su 25ga 3-1/2in quincke were found discolored before use.The following information was provided by the initial reporter: "reported issue: discoloration".
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary 63 samples were received by our quality team for evaluation.Upon visual inspection and evaluation of all 63 samples, it was observed that the samples were in product specification; therefore, the incident could not be verified.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.Based on the investigation, the samples did not contain discoloration, variation in color, foreign matter, nor cloudiness condition and meet product specification requirements.Since the incident could not be verified a root cause could not be determined.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
|
| |
|
Event or Problem Description
|
|
It was reported that 63 needles spinal s/su 25ga 3-1/2in quincke were found discolored before use.The following information was provided by the initial reporter: "reported issue: discoloration".
|
| |
|