| Model Number |
288304000 |
| Medical Device Problem Code |
Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Reporter is a synthes employee.Without a lot number, the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for, and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
Device report from (b)(6) reports an event as follows: it was reported that on an unknown date during the inspection of the mountaineer trays at the hospital, their poly axial screwdrivers were worn out and needed to be replaced.This report is for one (1) mini polyaxial screwdriver.This is report 5 of 6 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a review of the receiving inspection (ri) for minipolyaxial screwdriver was conducted identifying that lot number km868746 was released in three batches.Batch1: released on october 26, 2018 with no discrepancies.Batch2: released on october 26, 2018 with no discrepancies.Batch3: released on november 15, 2018 with no discrepancies.As a result, the ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Visual inspection: the instrument was received at us customer quality (cq).The distal tip is stripped, preventing the device from functioning as intended.The stripped condition of the device is consistent as an end of life indicator for the device.No other issues were identified during inspection.The received condition is consistent with the complaint condition thus the complaint is confirmed.Please note only the outer driver assembly and sleeve assembly was returned; the inner driver sleeve assembly was not returned.A device failure was identified.Investigation conclusion: after a visual inspection per guidance, it is determined that the reusable instrument is worn from repeated use and servicing; therefore, further investigation for the reported complaint device is not required.During the investigation, no product design or manufacturing issues were observed that may have contributed to the complaint condition; therefore, further corrective and/or preventive action is not required.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|