Model Number 283512
Device Problem Degraded (1153)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Investigation summary: the device was received and evaluated at the service center.The reported complaint that the device does not turn due to a jammed motor, was confirmed.It was found that the resistance values of the keypad and the protective earth resistance of the motor cable were out of range.The motor was found to be corroded and runs not constantly.The motor, motor cable and the hand control set were replaced and the device was repaired, tested and found to be fully functional.Fluid ingress into the system and contact with the motor is responsible for the corrosion of the motor and could have caused the damage to the motor cable and the keypad of the hand control set.The corroded motor has a tendency to stick and not turn, hence is responsible for the issue reported by the customer.A manufacturing record evaluation was performed for the finished device (serial number: (b)(4)), and no non-conformance's were identified.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Event Description
It was reported by the affiliate in (b)(6) that during an arthroscopy procedure on an unknown date, it was observed that the micro tornado hp w hand control device was jammed that it would not turn.During in-house engineering evaluation, it was determined that the motor was corroded.There was no delay in the surgical procedure.There was a spare device available for use to complete the surgery.There was patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6013142063
MDR Report Key10726744
MDR Text Key212730875
Report Number1221934-2020-03188
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeLO
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/22/2020
Date Manufacturer Received10/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-