| Model Number |
03P6832 |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
09/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A review of tickets determined that there is normal complaint activity for architect magnesium reagent lot number 57754un20.Trending review determined no trends for discrepant patient results for the product.World wide data from abbott link was reviewed and determined that the patient median result for magnesium reagent lot number 57754un20 is within established control limits.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of the architect magnesium reagent lot number 57754un20 was identified.Patient identifier complete sid: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported a falsely elevated magnesium result generated on the architect c4000 analyzer on one patient.Results provided: (b)(6) 2020; (b)(6)= 7.4 / 1.78 mg/dl.Normal range: 1.6-2.6 mg/dl.No impact to patient management was reported.
|
| |
|