• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT GMBH MAGNESIUM; PHOTOMETRIC METHOD, MAGNESIUM Back to Search Results
Model Number 03P6832
Medical Device Problem Code High Test Results (2457)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
A review of tickets determined that there is normal complaint activity for architect magnesium reagent lot number 57754un20.Trending review determined no trends for discrepant patient results for the product.World wide data from abbott link was reviewed and determined that the patient median result for magnesium reagent lot number 57754un20 is within established control limits.Return testing was not completed as returns were not available.A review of the manufacturing documentation did not identify any issues associated with the customer's observation.A review of the product labeling concluded that the issue is sufficiently addressed.Based on the investigation, no systemic issue or deficiency of the architect magnesium reagent lot number 57754un20 was identified.Patient identifier complete sid: (b)(6).
 
Event or Problem Description
The customer reported a falsely elevated magnesium result generated on the architect c4000 analyzer on one patient.Results provided: (b)(6) 2020; (b)(6)= 7.4 / 1.78 mg/dl.Normal range: 1.6-2.6 mg/dl.No impact to patient management was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGNESIUM
Common Device Name
PHOTOMETRIC METHOD, MAGNESIUM
Manufacturer (Section D)
ABBOTT GMBH
max-planck-ring 2
wiesbaden 65205
GM  65205
MDR Report Key10726763
Report Number3002809144-2020-01043
Device Sequence Number3457136
Product Code JGJ
UDI-Device Identifier00380740161637
UDI-Public00380740161637
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K181748
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/25/2021
Device Model Number03P6832
Device Catalogue Number03P68-32
Device Lot Number57754UN20
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 09/30/2020
Initial Report FDA Received Date10/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARC C4000 INTGR, 02P24-40, C402250
-
-