Device evaluated by mfr.: p elite ous mr 24 x 3.00 mm stent delivery system was returned for analysis.A visual examination of the stent found no issues.There was no sign of damage, stretching or lifting of the stent struts.The stent showed no signs of movement and was set between the proximal and distal markerbands.The crimped stent outer diameter was measured and the result was within maximum crimped stent profile measurement.The balloon cones were reviewed, and no issues were noted.The balloon wings were tightly wrapped and evenly folded and were not subjected to positive pressure.A visual and microscopic examination of the tip showed no signs of damage.A visual and tactile examination of the hypotube found multiple kinks along several locations of the hypotube shaft.A hypotube break was also noted at 840 mm distal to the distal end of the strain relief.This type of damage is consistent with excessive force that could have been applied to the delivery system.A visual and tactile examination of the outer and mid-shaft section and the visual inspection of the inner extrusion found no issues along the shaft polymer extrusion.No other issues were identified during the product analysis.
|
Reportable based on device analysis completed on 02-oct-2020.It was reported that crossing difficulties were encountered.Vascular access was obtained via the femoral artery.The 74% stenosed, 22x3.0, eccentric, de novo target lesion with a bend of >45 and <90 degrees was located in the moderately tortuous and moderately calcified left anterior descending artery.A 24 x 3.00 promus elite drug-eluting stent was advanced but failed to cross the lesion.The procedure was completed with another of the same device.There were no patient complications reported.Patient was stable.However, returned device analysis revealed shaft break.
|