| Model Number |
24700 |
| Medical Device Problem Codes |
Deflation Problem (1149); Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/09/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that a new access site was created in order to deflate the balloon.A 7.0mm x 40mm x 80cm 4f sterling was selected for a percutaneous transluminal angioplasty procedure in the occluded arteriovenous shunt vessel.During the procedure, the balloon failed to fully deflate.While attempting to withdraw, the sterling could not be removed through the 4f non-boston scientific introducer sheath.The patient was punctured again and a retraction cannula was used to pierce and deflate the balloon.The sterling was removed from the sheath and the procedure was successfully completed.No patient complications were reported.
|
| |
|
Event or Problem Description
|
|
It was reported that a new access site was created in order to deflate the balloon.A 7.0mm x 40mm x 80cm 4f sterling was selected for a percutaneous transluminal angioplasty procedure in the occluded arteriovenous shunt vessel.During the procedure, the balloon failed to fully deflate.While attempting to withdraw, the sterling could not be removed through the 4f non-boston scientific introducer sheath.The patient was punctured again and a retraction cannula was used to pierce and deflate the balloon.The sterling was removed from the sheath and the procedure was successfully completed.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by manufacturer: returned product consisted of a sterling balloon catheter.The shaft, hypotube, tip and balloon were microscopically and visually examined.There was contrast in the inflation lumen and balloon, and blood in the guidewire lumen.The balloon was loosely folded.There was a longitudinal tear 2mm long at the distal marker band consistent with the reported information that the balloon was punctured in order to deflate the device.There was no marker band damage detected.The outer shaft was stretched down for 10mm starting 58.1cm from the hub, and 6mm starting 73.3cm from the hub.The stretched down outer shaft would have prevented the balloon from being able to deflate.
|
| |
|
Search Alerts/Recalls
|