• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. PENUMBRA SMART COIL; HCG, KRD Back to Search Results
Model Number 400SMTXSFT1H04
Medical Device Problem Code Separation Failure (2547)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2020-01815.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the anterior communicating artery (acom) using penumbra smart coils (smart coils), a penumbra smart coil detachment handle (handle) and non-penumbra microcatheter.During the procedure, the physician advanced a smart coil into the target location using the microcatheter and attempted to detach it using the handle; however, the smart coil failed to detach.Therefore, the smart coil and handle were removed.The procedure was completed using a new smart coil and new handle.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pusher assembly was fractured on its proximal end.The pusher assembly was kinked approximately 4.0 and 8.5 cm from its proximal end.The pull wire was inside the distal detachment tip (ddt) and the embolization coil was intact with its pusher assembly and had offset coil winds.There was no visible damage to the braided shaft.Conclusions: evaluation of the returned handle revealed a functional device.During the functional test, the handle was functionally tested on a handle fixture and passed within specification.The handle was then used to detach a demonstration coil without an issue.Evaluation of the returned smart coil confirmed that the embolization coil was not detached from the pusher assembly.Further evaluation revealed that the proximal end of the pusher assembly was fractured off and the pull wire was inside the ddt.This damage was likely incidental to the reported complaint.However, this damage prevented the smart coil from being functionally tested.Therefore, the root cause of the reported issue could not be determined.Penumbra handles are functionally tested during incoming inspection by quality.Penumbra coils are visually inspected during in-process inspection and during quality inspection after manufacturing.Section h.Box 6.Conclusions code 1: 4316 - the investigation findings do not lead to a clear conclusion about the root cause of smart coil detachment failure.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report number: 1.3005168196-2020-01815.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA SMART COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key10727896
Report Number3005168196-2020-01816
Device Sequence Number3345574
Product Code HCG
UDI-Device Identifier00814548015248
UDI-Public00814548015248
Combination Product (Y/N)Y
Initial Reporter CountryPO
PMA/510(K) Number
K160832
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/02/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/27/2025
Device Model Number400SMTXSFT1H04
Device Catalogue Number400SMTXSFT1H04
Device Lot NumberF95910
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2020
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 10/02/2020
Supplement Date Received by Manufacturer11/24/2020
Initial Report FDA Received Date10/23/2020
Supplement Report FDA Received Date01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA01/01/2005
-
-