• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-55-040-120-P6
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned for evaluation.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no indication of a lot specific product quality issue.Based on the information provided, a conclusive cause for the reported difficulty could not be determined.It is possible that the introducer sheath was too close to the stent during deployment causing a portion of the stent to deploy within the introducer sheath; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the superficial femoral artery.Pre-dilatation was performed, and a 5.5x150mm supera self-expanding stent was implanted.Then a 5.5x40mm supera stent was deployed, but the stent remained in the introducer sheath and was removed.A non-abbott stent was used to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10727958
Report Number2024168-2020-08873
Device Sequence Number3312259
Product Code NIP
UDI-Device Identifier08717648226069
UDI-Public08717648226069
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2022
Device Model NumberS-55-040-120-P6
Device Catalogue NumberS-55-040-120-P6
Device Lot Number0040861
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/07/2020
Initial Report FDA Received Date10/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age44 YR
Patient Weight45
-
-