| Model Number |
NGP680703 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Vitrectomy (2643)
|
| Date of Event |
09/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Trouble-shooting with technical support (ts) was performed and operation of the foot pedal was tested remotely with the system test process.Ts verified that bluetooth communication and connectivity between footpedal and console was working properly.A field service engineer (fse) was dispatched to customer site.The fse reconnected the footpedal and confirmed that it passed full function and calibration tests.Battery was charged at 100 percent.A review of records related to the whitestar signature pro adv linear pedal including labeling, complaint trending, and risk documentation will be performed.Upon completion of this review, if any further relevant information is obtained, a supplemental medwatch will be filed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
Lost communication between the footpedal and phaco system resulting in an unplanned vitrectomy was reported.A brief description from the surgery center indicated that during cataract removal surgery on (b)(6) 2020, the footpedal and phaco machine lost communication during a dense cataract procedure of a male patient¿s right eye (od).Vitrectomy was required.The cataract surgery was completed using the same phaco console.This report is for the whitestar signature pro adv linear pedal.A separate report will be submitted for the whitestar signature pro adv linear console.
|
| |
|
Additional Manufacturer Narrative
|
|
H4: the correct manufacturing date is 10/17/2017 as provided in this report.All pertinent information available to johnson & johnson surgical vision, inc has been submitted.
|
| |
|
Search Alerts/Recalls
|