• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9438-06
Medical Device Problem Code Wireless Communication Problem (3283)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/27/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).
 
Event or Problem Description
It was reported that signal loss over one hour occurred.A review of the share logs was performed and signal loss was found within the investigation window.The allegation was confirmed.The probable cause was the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Common Device Name
CONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence dr.
san diego CA 92121
MDR Report Key10729968
Report Number3004753838-2020-124879
Device Sequence Number13883321
Product Code PQF
UDI-Device Identifier00386270000620
UDI-Public00386270000620
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P120005
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model Number9438-06
Device Catalogue NumberSTT-MC-001
Device Lot Number7266961
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 09/27/2020
Initial Report FDA Received Date10/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age58 YR
-
-