• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT VASCULAR SUPERA SELF-EXPANDING STENT SYSTEM; SELF EXPANDING PERIPHERAL STENT SYSTEM Back to Search Results
Model Number S-50-080-120-P6
Medical Device Problem Code Activation Failure (3270)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the procedure was to treat a de novo lesion from the superficial femoral artery to the p1 segment.A cross over approach from femoral was performed.A guide wire was crossed to the lesion, pre-dilatation was performed, and then a 5x80mm supera was attempted to deploy.The thumbslide was fully advanced, then the deployment lock was unlocked and the stent was deployed.When the stent delivery system was attempted to be removed under fluoroscopy, the stent implant moved proximal and a part of the stent moved into the sheath.It was decided to remove the entire system including the stent implant with the sheath.The sheath was changed and a new 5x100mm supera and non-abbott stent were implanted to successfully complete the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
Visual analysis was performed on the returned device.The reported difficulties were unable to be confirmed due to the condition of the returned device.A review of the lot history record revealed no manufacturing nonconformities that would have contributed to this complaint.Additionally, a review of the complaint history of the reported lot revealed no other complaint reported from this lot.The investigation was unable to determine a cause for the reported difficulties.It may be possible that the distal sheath of the delivery system was entrapped or bent in the anatomy such that the ratchet was unable to properly engage the stent during the final deployment stroke of the thumbslide, or possible that the tip caught on the stent during retraction causing the stent to pull back with the delivery system; however, this could not be confirmed.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.Na.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUPERA SELF-EXPANDING STENT SYSTEM
Common Device Name
SELF EXPANDING PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10730448
Report Number2024168-2020-08882
Device Sequence Number3312264
Product Code NIP
UDI-Device Identifier08717648211751
UDI-Public08717648211751
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
P120020
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2021
Device Model NumberS-50-080-120-P6
Device Catalogue NumberS-50-080-120-P6
Device Lot Number9091161
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/28/2020
Initial Date Received by Manufacturer 10/08/2020
Supplement Date Received by Manufacturer11/10/2020
Initial Report FDA Received Date10/23/2020
Supplement Report FDA Received Date11/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
GUIDING SHEATH, DESTINATION; GW, VASSALLO FLOPPY
-
-