ABBOTT VASCULAR XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM; DRUG ELUTING CORONARY STENT DELIVERY SYSTEM Back to Search Results
Model Number 1550225-08
Medical Device Problem Code Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no similar incidents from this lot.A review of the xience sierra label attached to the lot history record for this lot was conducted indicating a manufacturing date of 26-september-2019 with an expiration date (use by date) of (b)(6) 2020.The product was used after expiration as it was reported the procedure occurred on (b)(6) 2020.The expiration date of the product is important for sterility, efficacy, and performance of the device.It should be noted that the xience sierra, instructions for use states: note the product use by (expiration) date specified on the package.It is likely the ifu deviation of device expiration issue contributed to the event.The investigation determined the reported device expiration issue appears to be related to the use error.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was performed to treat a lesion in the mid second diagonal coronary artery.A 2.25x8mm xience sierra stent was implanted on (b)(6) 2020, and after the procedure, it was noted that the expiration date was (b)(6) 2020.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIENCE SIERRA EVEROLIMUS ELUTING CORONARY STENT SYSTEM
Common Device Name
DRUG ELUTING CORONARY STENT DELIVERY SYSTEM
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10730471
Report Number2024168-2020-08881
Device Sequence Number3312265
Product Code NIQ
UDI-Device Identifier08717648226991
UDI-Public08717648226991
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P110019
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/25/2020
Device Model Number1550225-08
Device Catalogue Number1550225-08
Device Lot Number9091941
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/08/2020
Initial Report FDA Received Date10/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-