Catalog Number 5111-00300-060
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Event or Problem Description
It was reported that a 3.0mm coupler device had one side of the wing jaw that was broken; further described as a broken tip.This was identified during preparation for surgery.There was no patient involvement.No additional information is available.
 
Additional Manufacturer Narrative
H10: the device was received for evaluation.A visual inspection was performed which observed the right jaw assembly was broken.The jaw was missing both the ring and half of the right jaw.The remaining coupler ring showed no signs of use.The reported condition was verified.The cause of the condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
COUPLER
Common Device Name
DEVICE, ANASTOMOTIC, MICROVASCULAR
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
MDR Report Key10730834
Report Number1416980-2020-06610
Device Sequence Number11627231
Product Code MVR
UDI-Device Identifier00844735002676
UDI-Public00844735002676
Combination Product (Y/N)Y
Initial Reporter CountryJA
PMA/510(K) Number
K861985
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 12/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number5111-00300-060
Device Lot NumberSP20C13-1436267
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/04/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer11/19/2020
Initial Report FDA Received Date10/23/2020
Supplement Report FDA Received Date12/01/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No