Model Number PCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643)
Date of Event 09/30/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted: if explanted: not applicable, as the intraocular lens was inserted and removed in the initial surgery.Device evaluation: product testing could not be performed because the product was not returned.A product quality deficiency could not be confirmed.The reported issue was not verified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.Historical data analysis: a search revealed that no additional complaints for this production order number have been received.Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported intraocular lens (iol) was implanted and explanted.Through follow-up, additional information was received stating iol was explanted (removed) after a rupture of the posterior capsular bag was noted during the removal of viscoelastic.Lens was explanted (removed) without difficulty and was not related to a product quality issue.A sulcus iol was implanted during the same procedure after anterior vitrectomy was performed.The patient is doing well.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS ITEC PRELOADED 1-PIECE IOL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e. st. andrew place
santa ana CA 92705
MDR Report Key10730982
Report Number2648035-2020-00772
Device Sequence Number1114987
Product Code HQL
UDI-Device Identifier05050474558366
UDI-Public(01)05050474558366(17)230728
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/28/2023
Device Model NumberPCB00
Device Catalogue NumberPCB0000250
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/01/2020
Initial Report FDA Received Date10/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age71 YR
-
-