| Model Number |
PCB00 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Capsular Bag Tear (2639); Vitrectomy (2643)
|
| Date of Event |
09/30/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
If implanted: if explanted: not applicable, as the intraocular lens was inserted and removed in the initial surgery.Device evaluation: product testing could not be performed because the product was not returned.A product quality deficiency could not be confirmed.The reported issue was not verified.Manufacturing records review: the manufacturing records for the product were reviewed, the product was manufactured and released according to specification.Historical data analysis: a search revealed that no additional complaints for this production order number have been received.Conclusion: as a result, of the investigation there is no indication of a product malfunction or product quality deficiency.The reported issue was not verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported intraocular lens (iol) was implanted and explanted.Through follow-up, additional information was received stating iol was explanted (removed) after a rupture of the posterior capsular bag was noted during the removal of viscoelastic.Lens was explanted (removed) without difficulty and was not related to a product quality issue.A sulcus iol was implanted during the same procedure after anterior vitrectomy was performed.The patient is doing well.No further information is available.
|
| |
|