• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE ICONIX 1 TT WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE ICONIX 1 TT WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 3910-500-312
Medical Device Problem Code Material Fragmentation (1261)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that the inserter of the device is potentially what broke.All pieces were retrieved.
 
Event or Problem Description
It was reported that the inserter of the device is potentially what broke.All pieces were retrieved.
 
Additional Manufacturer Narrative
This complaint investigation was closed based on the device not received, therefore the reported failure mode was not confirmed.In the event that the device is received, the complaint will be reopened and the investigation will be updated with new results.Alleged failure: inserter or the anchor breakage.Probable root cause: design; inserter/implant not compatible with guide; inserter material/design too weak; inserter shaft cannot bend around curved guide; drill not designed to center hole with respect to guide; guide mouth not designed to grip bone; guide not able to be gripped tightly.Process: inserter, implant, and/or guide manufactured out of spec; anchor coating process out of specification.Application: excessive force impacting inserter; movement of guide out of orientation to pilot hole; use of wrong drill.The reported failure mode will be monitored for future reoccurrence.Manufacture date is not known.H3 other text : 81.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ICONIX 1 TT WITH INTELLIBRAID TECHNOLOGY, 1.4MM ANCHOR WITH 1.2MM XBRAID TT
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10731147
Report Number0002936485-2020-00427
Device Sequence Number11670131
Product Code MBI
UDI-Device Identifier37613327176927
UDI-Public37613327176927
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K133671
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number3910-500-312
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/25/2020
Supplement Date Received by Manufacturer09/25/2020
Initial Report FDA Received Date10/23/2020
Supplement Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-