• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC HEART VALVES DIVISION MOSAIC AORTIC BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC HEART VALVES DIVISION MOSAIC AORTIC BIOPROSTHETIC HEART VALVE; HEART-VALVE, NON-ALLOGRAFT TISSUE Back to Search Results
Model Number 305U219
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Insufficient Device Problem Information (3190)
Health Effect - Clinical Codes Aortic Valve Stenosis (1717); No Information (3190)
Date of Event 05/14/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product analysis: no product was returned.Conclusion: without the return of the product, no definitive conclusion can be made regarding the clinical observation.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Medtronic received information from the patients daughter that 5 years and 5 months post implant of this 19 mm aortic bioprosthetic valve, it was explanted and replaced with a non medtronic product.The reason for the replacement was not reported.No additional adverse patient effects were reported.
 
Additional Manufacturer Narrative
Medtronic received additional information that the reason for replacement was progressive aortic stenosis with severe patient-prosth esis mismatch.It was reported that the valve leaflets appeared completely normal, there was "simply valvular dysfunction related to the size of the valve".The 19mm valve was replaced with a 23mm non-medtronic valve.No additional adverse patient effects were r eported.A4: patient weight added b7: relevant history added g: manufacturer contact updated h6: coding updated.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOSAIC AORTIC BIOPROSTHETIC HEART VALVE
Common Device Name
HEART-VALVE, NON-ALLOGRAFT TISSUE
Manufacturer (Section D)
MEDTRONIC HEART VALVES DIVISION
1851 e deere ave
santa ana CA 92705
MDR Report Key10731268
Report Number2025587-2020-03222
Device Sequence Number1872311
Product Code DYE
UDI-Device Identifier00643169001879
UDI-Public00643169001879
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P990064
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2013
Device Explanted Year2019
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/08/2016
Device Model Number305U219
Device Catalogue Number305U219
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 09/28/2020
Supplement Date Received by Manufacturer10/30/2020
Initial Report FDA Received Date10/23/2020
Supplement Report FDA Received Date11/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
Patient Age78 YR
Patient Weight79
-
-