Model Number AAB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 09/29/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If implanted, give date: not applicable, as lens was removed/replaced during the same procedure.If explanted, give date: not applicable, as lens was removed/replaced during the same procedure.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) was implanted and removed in the same day as the patient's capsule tore.A vitrectomy, incision enlargement and sutures were required.Another johnson & johnson lens (model ar40e 21.0 diopter) was successfully implanted in the sulcus as a replacement.No additional information was provided to johnson & johnson surgical vision.
 
Additional Manufacturer Narrative
Additional information: section d10: device available for evaluation? yes.Returned to manufacturer on: 10/21/2020.Section h3: device returned to manufacturer ¿ yes.Device evaluation: the complaint lens was received in the original lens case.The original folding carton, patient stickers, patient id card, customer notes (post-it), and implant notification card were received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half (but not separated), which is consistent with a lens handled during removal and replacement.Based on the return condition of the lens, no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no other complaints for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSAR
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10731272
Report Number2648035-2020-00773
Device Sequence Number1114988
Product Code HQL
UDI-Device Identifier05050474548695
UDI-Public(01)05050474548695(17)240318
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/18/2024
Device Model NumberAAB00
Device Catalogue NumberAAB0000215
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer11/10/2020
Initial Report FDA Received Date10/23/2020
Supplement Report FDA Received Date12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-