| Model Number |
15-752 |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
08/26/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade ii/iii.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported right side capsular contracture baker grade ii/iii.Device remains implanted.
|
| |
|
Event or Problem Description
|
|
Device has been explanted.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified the weight the device within specification, deformation, voids, and wear abrasion.A cloudy color in the gel was observed after the autoclave cycle.Based on the device analysis, the final assessment is no issues found related with the manufacturing process.
|
| |
|
Search Alerts/Recalls
|