Allergan did not submit this mdr within 30 days of becoming aware.Recent stimulated reporting related to 2011068-7/2/19-0101-r has increased complaint and mdr volume.Allergan is implementing a plan to address the increased volumes.A review of the device history record has been completed.No deviations or non-conformances noted.Device evaluation: visual analysis of the returned device identified: opening, white particles, wear abrasion, crease flat and weight to the spec.A microscopic analysis was performed which identify puncture opening on anterior side, consist in the use of some surgical tool.Based on the device analysis the final assessment is: two puncture opening on anterior due to surgical damage like.The event of capsular contracture is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture baker grade iii.
|