| Model Number |
0064599 |
| Medical Device Problem Code |
False Negative Result (1225)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional serial numbers: (b)(4).Additional lot numbers: n602, ua16765897, ua 15757557, 630060.
|
| |
|
Event or Problem Description
|
|
This report summarizes 16 malfunction events.A review of the events indicated that fourteen (14) patient samples tested on the galileo neo, automated blood bank system produced inaccurate results.A review indicated that three (3) patient sample tests using the galileo neo automated blood bank system produced an unexpected false negative result for the anti-e antibody; one (1) for the anti-c antibody; one (1) for the anti-c antibody; one (1) for the anti-jka antibody; one (1) for the anti-jkb antibody; two (2) for the anti-k antibody; one (1) for the anti-s antibody and one (1) for the anti-m antibody and one (1) for anti-d antibody.Four (4) instrument malfunctions produced inaccurate abo using the abo forward typing assay based on historical results for the patients.Subsequent testing of reagent retention lots, reviews of design history records and reagent antigen validation testing of the initial bulk product used for commercial vialing showed acceptable results.Through post event tests and investigations, no specific causes were determined for the malfunctions.
|
| |
|