• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TFCC MENDER DISPOSABLE SUTURE SYSTEM; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TFCC MENDER DISPOSABLE SUTURE SYSTEM; PASSER Back to Search Results
Model Number 7210752
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 09/25/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during arthroscopy procedure, the tfcc repair break away from the hubs when trying to remove from the plastic tray they are packaged in.No significant delay, and a back up was available to complete the procedure.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Additional Manufacturer Narrative
The reported device was received for evaluation.A visual inspection of the returned device found that it is not in its original packaging.Two suture retrievers were returned, the hubs have fractured away from the shaft and snare of both retrievers.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found no similar reported events.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.The root cause was associated with device design.A design change was implemented as a result of a corrective action initiated to address the reported issue.A stress relief collar is now attached at the connection between the shaft and the hub of the suture capture loop.Further corrective actions are not recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFCC MENDER DISPOSABLE SUTURE SYSTEM
Common Device Name
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10732165
Report Number1219602-2020-01654
Device Sequence Number1901990
Product Code HWQ
UDI-Device Identifier03596010563279
UDI-Public03596010563279
Combination Product (Y/N)N
Initial Reporter CountryCA
PMA/510(K) Number
K885311
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source Other,Foreign,User Facility,Company Representative
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2024
Device Model Number7210752
Device Catalogue Number7210752
Device Lot Number2046149
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Received by Manufacturer 09/30/2020
Supplement Date Received by Manufacturer09/30/2020
02/29/2024
Initial Report FDA Received Date10/25/2020
Supplement Report FDA Received Date12/04/2020
03/04/2024
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured12/06/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient SexUnknown
-
-