Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime or a33 test, excessive no delivery test and delivery accuracy test at 0.0873 inches.The stop (idle) current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device also passed tone check and vibrate check during self test.No audio or vibrate or absence of alarm anomalies noted during testing.The no delivery alarm functions properly during the basic occlusion test and occlusion test.No unexpected no delivery alarm noted during testing.Device uploaded properly using carelink.Device had cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.(b)(4).
|