CONCORD MANUFACTURING 2008T GEN 2 BIBAG WITHOUT CDX; DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER) Back to Search Results
Model Number 190895
Medical Device Problem Codes Thermal Decomposition of Device (1071); Melted (1385)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 10/07/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A biomedical technician (biomed) at a user facility reported that a fresenius 2008t hemodialysis (hd) machine had a burnt power plug.A patient was not connected to the machine at the time of the incident and there was no harm to any patients or individuals because of this malfunction.The biomed reported that the power plug charred and melted and made a burning smell, however there was no smoke, spark, flame, or arcing observed.The biomed reported that the machine has not had any past problems with failing the electrical leakage test.The machine is plugged into a hospital grade ground-fault circuit interrupter (gfci) outlet.The power cord was replaced, but the machine remained unable to power on.The biomed then replaced the power supply, at which point the machine was able to power on.The power supply did not appear burned, charred, melted, or black.There was no other damage observed on any other components.The biomed reported that the machine has been returned to service without issue.The sample was reported to be unavailable to be returned to the manufacturer for physical evaluation.
 
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation.Additionally, no on-site evaluation was performed by a fresenius regional equipment specialist (res).A records review was performed on the reported serial number.An investigation of the device manufacturing records was conducted by the manufacturer.There were no non-conformances, or any associated rework identified during the manufacturing process which could be related to the reported event.In addition, the device history record (dhr) review confirmed the results of the in-progress and final quality control (qc) testing met all requirements.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T GEN 2 BIBAG WITHOUT CDX
Common Device Name
DIALYSATE CONCENTRATE FOR HEMODIALYSIS (LIQUID OR POWDER)
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
MDR Report Key10733116
Report Number2937457-2020-01928
Device Sequence Number14238103
Product Code KPO
UDI-Device Identifier00840861100927
UDI-Public00840861100927
Combination Product (Y/N)N
Initial Reporter StateGA
Initial Reporter CountryUS
PMA/510(K) Number
K121341
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 11/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number190895
Device Catalogue Number190895
Was Device Available for Evaluation? No
Device AgeMO
Initial Date Received by Manufacturer 10/07/2020
Supplement Date Received by Manufacturer10/29/2020
Initial Report FDA Received Date10/26/2020
Supplement Report FDA Received Date11/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-