Medtronic received a report that the 5.0x12 pipeline device was introduced to the vessel distal to the aneurysm.However, the distal portion of the stent would not open after several attempts.The stent was removed, and during retrieval, the delivery pushwire separated from the implant near the hub of the phenom 27 microcatheter.The stent was removed along with the catheter, and replacement products were used to complete the procedure.It was noted the physician used significant force to remove the pipeline due to resistance in the hub of the catheter, and was unknown if there was any damage to the catheter.It was indicated that all devices were prepared as per the instructions for use (ifu), and no patient symptoms or further complications were reported as a result of this event.Post-procedure angiography showed an implanted pipeline with good apposition and coverage of the neck of the aneurysm.The patient was undergoing surgery for treatment of a saccular, unruptured aneurysm of the left cavernous segment of the internal ca rotid artery with a max diameter of 12 mm and a 5 mm neck diameter.It was noted the patient's vessel tortuosity was moderate.Dual antiplatelet therapy (dapt) was administered, and the platelet reactivity units (pru) level was unknown.Ancillary devices include a neuron max sheath, phenom plus guide catheter, phenom 27 microcatheter.
|