| Model Number |
L20106 |
| Medical Device Problem Code |
Loss of Osseointegration (2408)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646); No Code Available (3191)
|
| Date of Event |
10/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that the patient was revised due to pain.Femoral stem was slightly loose and removed.Revised with a biomet taper loc stem.Doi: unknown (approximately 11 years ago).Dor: (b)(6) 2020; right hip.
|
| |
|
Additional Manufacturer Narrative
|
|
Product complaint # (b)(4).Investigation summary = no device associated with this report was received for examination.Three provided patient x-rays have been reviewed.It was not possible to confirm a loosening event with the provided images.Nothing indicative of a product problem is identified.The information received will be retained for potential series investigations if triggered by trend analysis or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
|
| |
|
Search Alerts/Recalls
|