Additional Manufacturer Narrative
|
|
This device was returned for service, however, did not meet manufacturing specifications during pre-repair assessment.During repair, it was determined that the device had cutter whip, and did not function.Therefore, the reported condition was confirmed.The assignable root cause of this condition was determined to be traced to component failure due to normal wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from the (b)(6) that during service and evaluation, it was determined that the coupling device had cutter whip, and did not function.It was further determined that the device failed pretest for function test.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention, or prolonged hospitalization.The exact date of the event was not reported, however, it was reported that the event occurred in 2020.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|