| Model Number |
1011-0139-01 |
| Medical Device Problem Code |
Failure to Deliver Shock/Stimulation (1133)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has not received the device for evaluation and this complaint is still under investigation.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that while attempting to treat a patient (age & gender unknown), the associated defibrillator failed to discharge using these internal handles.Complainant indicated that there was no adverse effect to the patient due to the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
Zoll medical canada evaluated the internal handles and the customer's report was not replicated or confirmed.The internal handles were put through extensive testing including full functional testing and multiple shock testing without duplicating the report.The internal handles were returned to the customer.No trend is associated with reports of this type.An internal inspection resulted in no findings.
|
| |
|