TERUMO CARDIOVASCULAR SYSTEMS CORPORATION STERILE FX25REC W/RES; BLOOD GAS OXYGENATOR Back to Search Results
Model Number 3CX*FX25REC
Device Problem Defective Component (2292)
Patient Problem No Patient Involvement (2645)
Event Date 09/30/2020
Event Type  malfunction  
Manufacturer Narrative
Terumo has received the device for evaluation, however, the investigation has yet to be completed.Terumo plans on submitting a follow-up report when the investigation is complete and when more information becomes available.(b)(4).
 
Event Description
The user facility reported to terumo cardiovascular during other non-clinical activity, the ¿o¿ port top of oxy not fully formed.No patient involvement.
 
Manufacturer Narrative
This follow-up report is submitted to fda in accord with applicable regulations ¿ and as indicated by terumo cardiovascular systems in the initial report submitted to the fda on october 26, 2020.Upon further investigation of the reported event, the following information is new and/or changed: d4 (additional device information - added expiration date).G4 (date received by manufacturer).G7 (indication that this is a follow-up report).H2 (follow-up due to additional information).H3 (device evaluated by manufacturer).H4 (device manufacture date).H6 (identification of evaluation codes 10, 11, 3331, 3259, 4307).Method code #1: 10 - testing of actual/suspected device.Method code #2: 11 - testing of device from same lot/batch retained by manufacturer.Method code #3: 3331 - analysis of production records.Results code: 3259 - improper physical structure.Conclusions code: 4307 - cause traced to component failure.Upon review of the complaint sample the port was found to be damaged.A representative retention sample from the same lot number was inspected and found to be free of damage.All capiox units are 100% visually inspected at several points in the production process.All available information has been placed on file in quality management for appropriate tracking, trending, and follow-up.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STERILE FX25REC W/RES
Type of Device
BLOOD GAS OXYGENATOR
Manufacturer (Section D)
TERUMO CARDIOVASCULAR SYSTEMS CORPORATION
125 blue ball road
elkton MD 21921
MDR Report Key10737240
MDR Text Key213231655
Report Number1124841-2020-00243
Device Sequence Number1
Product Code DTZ
UDI-Device Identifier00699753450820
UDI-Public(01)00699753450820
Combination Product (y/n)N
PMA/PMN Number
K151791
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 11/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date02/28/2023
Device Model Number3CX*FX25REC
Device Catalogue NumberN/A
Device Lot NumberYD30
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/13/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/06/2020
Initial Date FDA Received10/26/2020
Supplement Dates Manufacturer Received11/18/2020
Supplement Dates FDA Received11/25/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-