Model Number 72200829
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Information (3190)
Date of Event 10/06/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, during surgery, at the moment the "6mm high flow cannula" was inserted in the point, the surgical staff realized that the metal part in the cannula was through the coagulation.They identified that was loose.The procedure was completed without delay using a back-up device.No additional complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Additional Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint facility for independent evaluation, thus visual inspection and functional testing could not be performed.A relationship, if any, between the subject device and the reported event could not be determined.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.The cann h.F.Dgnst 6mm rtbl dbl vlv instructions for use warns, ¿prior to use, inspect the device to ensure it is not damaged.Do not use a damaged device.¿ a review of the cann h.F.Dgnst 6mm rtbl dbl vlv risk management files was performed and found multiple line items addressing this failure mode.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.Internal complaint reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CANN H.F. DGNST 6MM RTBL DBL VLV
Common Device Name
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10737739
Report Number1219602-2020-01660
Device Sequence Number1767857
Product Code NBH
UDI-Device Identifier03596010573520
UDI-Public03596010573520
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number72200829
Device Catalogue Number72200829
Device Lot Number50557876
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer03/31/2021
Initial Report FDA Received Date10/26/2020
Supplement Report FDA Received Date04/06/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-