MAUDE Adverse Event Report: MPRI ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number 2AF284
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Cardiac Tamponade (2226); Pericardial Effusion (3271)
Date of Event 10/06/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during a cryo ablation procedure, the patient developed a cardiac tamponade with pericardial effusion.The case was aborted while the patient was under general anesthesia.The effusion was drained and the patient hospitalized for an extra day for monitoring.No further patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the 2af284 balloon catheter with lot number 00768 and the data files were returned and analyzed.Visual inspection of the balloon catheter showed the device was intact with no apparent issues.The smart chip verification indicated the catheter was used for five applications on the event date.The catheter passed the performance test and electrical integrity test as per specification; the impedance was also within specification.In conclusion, the reported pericardial effusion, tamponade, and aborting of the case with the patient under general anesthesia occurred during the procedure.There is no indication of a relation of the adverse event to the performance and malfunction of the product.The balloon catheter passed the returned product inspection as per specification.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE CARDIAC CRYOABLATION CATHETER
Common Device Name
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key10738544
Report Number2649622-2020-20688
Device Sequence Number1191546
Product Code OAE
UDI-Device Identifier00643169878365
UDI-Public00643169878365
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
P100010/S015
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 11/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/15/2021
Device Model Number2AF284
Device Catalogue Number2AF284
Device Lot Number00768
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/16/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer10/27/2020
Initial Report FDA Received Date10/26/2020
Supplement Report FDA Received Date11/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
990063-020 MAPPING CATHETER, 4FC12 SHEATH; 990063-020 MAPPING CATHETER, 4FC12 SHEATH
Outcome Attributed to Adverse Event Hospitalization; Life Threatening; Required Intervention;
-
-