| Model Number |
KD-V631M-07202S |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/10/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The subject device was returned to olympus medical systems corp.(omsc).The factory of aomori olympus confirmed that the cutting wire of the subject device was broken.The broken section of the cutting wire was melted and burned.As a measurement result of the outer diameter of the cutting wire, there was no abnormality.There were no missing parts.The cutting wire did not pierce the tube.The manufacturing record was reviewed and found no irregularities.It was known that the temperature of the cutting wire increased and the cutting wire broke if the device was activated under the following conditions where the distance between the cutting wire and the tissue was too close or the cutting wire contacted the tissue with a point contact.The above device handling has warned in the instruction.
|
| |
|
Event or Problem Description
|
|
We received the following report.During endoscopic sphincterotomy, the needle pierced the sheath.The user was unable to pass the device down an endoscope.No patient injury reported.The subject device was returned to olympus medical systems corp.(omsc) for evaluation.On (b)(6) 2020, olympus medical systems corp.(omsc) found that the cutting wire was broken off, and that the broken section of the cutting wire was melted and burned.
|
| |
|