Model Number 03.100.045
Medical Device Problem Code Device-Device Incompatibility (2919)
Health Effect - Clinical Code No Patient Involvement (2645)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Reporter is a synthes employee.The device was received, the investigation could not be completed, and no conclusion could be drawn, as product is entering the complaint system.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the four (4) stardrive screwdriver shafts were damaged.The stardrive shafts would not load into a standard ao quick connect chuck on the drill, several chucks and variations were tried.The issue was discovered during routine inspection.There was no patient involvement.Concomitant device reported: unknown ao quick connect (part# unknown, lot# unknown, quantity unknown).This report is for one (1) stardrive screwdriver shaft t15 250mm long.This is report 3 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.H3, h4, h6: part 03.100.045, lot 5446943: release to warehouse date: april 20, 2007.Manufactured by synthes monument.No non-conformance reports (ncrs) were generated during production.Review of the device history record showed that there were no issues during the manufacture of this product, and any subcomponents, which would contribute to this complaint condition.H3, h6: a product investigation was completed: upon visual inspection, the screwdriver shaft had a nick at the proximal end of the device and also scratches and scuffs consistent with field usage were observed on the device.No other issues were identified with the returned device.A functional test cannot be performed as the device was returned by itself.The diameter of the shaft was measured; it measured within the specified dimensions.The relevant documents were reviewed; no design issues or discrepancies were identified.The complaint condition cannot be confirmed.There is no indication that a design or manufacturing issue has caused the complaint condition and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STARDRIVE SCREWDRIVER SHAFT T15 250MM LONG
Common Device Name
SCREWDRIVER
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10739998
Report Number2939274-2020-04827
Device Sequence Number3141496
Product Code HXX
UDI-Device Identifier10886982070944
UDI-Public10886982070944
Combination Product (Y/N)N
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number03.100.045
Device Catalogue Number03.100.045
Device Lot Number5446943
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/14/2020
Initial Date Received by Manufacturer 09/30/2020
Supplement Date Received by Manufacturer10/29/2020
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date11/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN AO QUICK CONNECT; UNKNOWN AO QUICK CONNECT
-
-