C.R. BARD, INC. (COVINGTON) -1018233 BARD PLATINUM CLASS II FLAT WIRE STONE BASKET; BARD FLATWIRE STONE BASKETS Back to Search Results
Model Number 039690
Medical Device Problem Codes Material Fragmentation (1261); Misassembly by Users (3133)
Health Effect - Clinical Codes No Consequences Or Impact To Patient (2199); No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 09/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event or Problem Description
It was reported that one dormia probe was broken when the box was opened and basket rod not attached to the handle.For the second probe, the rod had detached from the handle after a single use and without even forcing it.Both were from the same lot.
 
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
 
Event or Problem Description
It was reported that one dormia probe broken when the box was opened and basket rod not attached to the handle.For the second probe, the rod had detached from the handle after a single use and without even forcing it.Both were from the same lot.
 
Event or Problem Description
It was reported that one dormia probe was broken when the box was opened and the basket rod not attached to the handle.For the second probe, the rod had detached from the handle after a single use without even forcing it.Both were from the same lot.
 
Additional Manufacturer Narrative
The reported event was confirmed.A stone basket was received in its original packaging with no retainer.The local knob was found to be loosened and the basket was removed from the handle, but was still in the sheath.The basket was able to be reassembled to manufacturing specifications and functioned properly.According the supplier, the user loosened the lock knob.In the original manufacturing condition the lock knob is tightened, and basket tip is the only portion outside of the sheath.It could not be properly packaged if the lock knob is loosened.A potential root cause for failure mode " detached basket rod" will be added.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "only physicians trained in stone manipulation should perform this procedure.A variety of techniques in the use of this instrument may be employed; however, the physician should use the technique with which he/she is most familiar.1.Inspect the device prior to use and during the procedure.Handle disassembly/reassembly: if handle removal is desired: 1.Unscrew the thumb screw located on the basket handle.2.Gently pull backward on the handle, releasing both the handle and sheath from the drivewire/basket combination.If any resistance is felt at this stage, stop and determine cause of resistance.3.Once the stone burden within the basket has been removed, the basket can be closed using an open tip ureteral catheter.4.The basket can be reassembled easily by inserting the drivewire into the sheath and advancing until it reaches the handle.5.Tighten the thumb screw to secure the drivewire in place." h11: section a through f - the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD PLATINUM CLASS II FLAT WIRE STONE BASKET
Common Device Name
BARD FLATWIRE STONE BASKETS
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10740381
Report Number1018233-2020-20453
Device Sequence Number3061333
Product Code FFL
UDI-Device Identifier00801741076558
UDI-Public(01)00801741076558
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 04/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/01/2022
Device Model Number039690
Device Catalogue Number045290
Device Lot NumberBMBUMM01
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Initial Date Received by Manufacturer 09/30/2020
Supplement Date Received by Manufacturer11/24/2020
04/05/2021
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date12/14/2020
04/27/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-