ZIMMER BIOMET, INC. TPRLC 133 T1 PPS HO 11X142MM 2MM T1; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Medical Device Problem Codes Device Contaminated During Manufacture or Shipping (2969); Packaging Problem (3007)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 07/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Report source: (b)(6).Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2020 - 03906, 0001825034 - 2020 - 03908.Reported event was confirmed by visual examination of device.Device history record (dhr) was reviewed and no discrepancies related to the reported event were found.The root cause of the foreign debris is the operator not following the work instructions provided.The root cause of the damaged blister is likely due to transit damage.Evaluation of the returned product/photographs provided confirmed the sterile packaging blister is damaged.Further evaluation found foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed.The sterility has been breached.The likely condition of the device when it left zimmer biomet is non-conforming to specification.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
It was reported that during an investigation of circulated items.Several devices were identified as having sterile packages damaged.Upon examination of device for reported packaging damage, it was discovered that the sterile packaging was not compromised but contained a foreign debris.No patients were involved.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS HO 11X142MM 2MM T1
Common Device Name
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10741843
Report Number0001825034-2020-03907
Device Sequence Number3456795
Product Code KWA
UDI-Device Identifier00880304489714
UDI-Public00880304489714
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K101086
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number51-104110
Device Lot Number6105011
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2020
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 10/23/2020
Initial Report FDA Received Date10/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-