Model Number 3501410
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 10/01/2016
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The mentor failure analysis lab has received the device for evaluation.The analysis has begun but is not complete at this time.When the investigational analysis has been completed, a supplemental report will be submitted.A manufacturing record evaluation (mre) was performed, and no anomalies were found related to this complaint.In addition, the mre verifies that the device was manufactured in accordance with documented specification and procedures.Reason for device explant and/or reoperation: deflation.(b)(4).
 
Event or Problem Description
It was reported that a caucasian female patient underwent a breast augmentation revision with a mentor smooth round spectrum 125cc saline breast prosthesis and experienced a deflation on the left side post-operatively, which was confirmed by a physician.As a result, the patient underwent explantation on (b)(6) 2020.
 
Additional Manufacturer Narrative
On november 24, 2020, mentor completed evaluation of the device.Device evaluation summary: during the visual evaluation, an anomaly was observed on the anterior view of the device consistent with a crease/fold.A tear was also observed within the crease/fold, measuring approximately 0.3 cm.The evaluation determined that the loss of shell integrity is consistent with a crease fold deflation of the implant shell, which is a known inherent risk of saline-filled mammary prosthesis.Deflation can occur at any time after implantation, but it is more likely to occur the longer the implant is implanted.Creating wrinkles or folds should be avoided during implantation or other procedures as it can result in a deflation in a later time.Breast implants may also simply wear out over time.A second product was received (product code 3501410 and lot number 5949064).No adverse events were reported for this concomitant (contralateral) device, therefore no further analysis is required.Each device is visually inspected during manufacturing to ensure the device meets the required specifications prior to shipment.Deflation complaint information is consistently analyzed and monitored by mentor quality assurance to determine when further action is necessary.Manufacturer¿s reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MENTOR SMOOTH ROUND SPECTRUM
Common Device Name
PROSTHESIS, BREAST, INFLATABLE, INTERNAL, SALINE
Manufacturer (Section D)
MENTOR TEXAS
3041 skyway circle north
irving TX 75038
MDR Report Key10742001
Report Number1645337-2020-13712
Device Sequence Number16925322
Product Code FWM
UDI-Device Identifier00081317001959
UDI-Public00081317001959
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P990075
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2010
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2013
Device Model Number3501410
Device Catalogue Number3501410
Device Lot Number5949064
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Initial Date Received by Manufacturer 10/06/2020
Supplement Date Received by Manufacturer11/24/2020
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date11/26/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age53 YR
-
-