| Catalog Number |
515053 |
| Medical Device Problem Codes |
Break (1069); Leak/Splash (1354)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
10/05/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the bd phaseal¿ optima injector (n35-o multi) "fell apart" when connected with the protector during use.The following information was provided by the initial reporter: "after drug extraction the injector n35-0 was falling apart when connected with the protector.Membrane and spring was removed from the shield and one piece of the injector stayed on the protector"."yes the product was used for chemotherapy drug ¿ perjeta.".
|
| |
|
Additional Manufacturer Narrative
|
|
H6: investigation summary two photos were provided to our quality team for investigation.Through visual inspection, the injector grips are observed to be separated, exposing the internal components.There is no visible damage noted in the photos on any of the components that may have contributed to this incident.Fifteen retained samples of lot 2004102 were used for additional evaluation.The product was visually inspected, no damage or defects were observed and all components were properly assembled.Functional testing was performed, connecting the injectors to a sample syringe, protector and vial according to the instructions for use.In all cases the product functioned properly, liquid inside the syringe could move to the vial and back to the syringe without issue, and no issues occurred during the connection or disconnection of the devices.It is possible that a misalignment of the injector grips during assembly resulted in the improper force being applied to join the components, therefore resulting in only partial assembly and the pieces disconnecting as observed in the product reported.A device history review was performed for lot 2004102, no deviations or non-conformances were identified during the manufacturing process that could have contributed to this failure.Product undergoes visual and functionals inspections throughout the manufacturing.Testing results were reviewed for the reported lot, no issues related to this failure were identified, all product met required specification.Based on the available information we are not able to identify a definitive root cause related to the manufacturing process at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that the bd phaseal¿ optima injector (n35-o multi) "fell apart" when connected with the protector during use.The following information was provided by the initial reporter: "after drug extraction the injector n35-0 was falling apart when connected with the protector.Membrane and spring was removed from the shield and one piece of the injector stayed on the protector" "yes the product was used for chemotherapy drug ¿ perjeta.".
|
| |
|