BOSTON SCIENTIFIC CORPORATION GLADIATOR ELITE; CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL Back to Search Results
Model Number 24686
Medical Device Problem Codes Leak/Splash (1354); Material Rupture (1546)
Health Effect - Clinical Code Hematoma (1884)
Date of Event 06/05/2020
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that balloon pinhole occurred and the patient developed a hematoma.The target lesion was located in an upper arm vessel.A 5.0 x40, 40cm gladiator elite balloon catheter was advanced for dilatation.However, during the first inflation at 6 atmospheres, it was noticed that the balloon was leaking fluid.The device was removed and tested for pressure.It was found that there was a hole in the balloon.The long operation time resulted in the patient developing a hematoma in the upper arm.The patient was transferred to a different hospital and sent to surgery.No further patient complications were reported.
 
Additional Manufacturer Narrative
Device evaluation by mfr: gladiator device - 5x40x40cm, 24atm was received for analysis.A visual examination identified that the balloon was not folded which indicates that the device was subjected to positive pressure.The returned device was attached to an encore inflation unit.Positive pressure was applied when liquid was observed to be leaking from a balloon pinhole located approximately 1mm proximal of the proximal markerband.An examination of the balloon material and markerbands identified no issues which could potentially have contributed to this complaint.A visual and microscopic examination of the markerbands identified no issues which may potentially have contributed to the complaint incident.A visual and tactile examination found no kinks or damage to the shaft of the device.A visual and tactile examination identified no issues or damage to the tip of the device.No issues were identified during the product analysis.
 
Event or Problem Description
It was reported that balloon pinhole occurred and the patient developed a hematoma.The target lesion was located in an upper arm vessel.A 5.0 x40, 40cm gladiator elite balloon catheter was advanced for dilatation.However, during the first inflation at 6 atmospheres, it was noticed that the balloon was leaking fluid.The device was removed and tested for pressure.It was found that there was a hole in the balloon.The long operation time resulted in the patient developing a hematoma in the upper arm.The patient was transferred to a different hospital and sent to surgery.No further patient complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLADIATOR ELITE
Common Device Name
CATHETER, ANGIOPLASTY, PERIPHERAL, TRANSLUMINAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key10744073
Report Number2134265-2020-14763
Device Sequence Number3141318
Product Code LIT
UDI-Device Identifier08714729809661
UDI-Public08714729809661
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K132810
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/22/2022
Device Model Number24686
Device Catalogue Number24686
Device Lot Number0024315302
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/01/2020
Initial Date Received by Manufacturer 10/13/2020
Supplement Date Received by Manufacturer12/03/2020
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date12/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age56 YR
Patient Weight61
-
-