ABBOTT VASCULAR NC TREK CORONARY DILATATION CATHETER Back to Search Results
Model Number 1012449-20
Medical Device Problem Codes Material Rupture (1546); Improper or Incorrect Procedure or Method (2017)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 10/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).The device is expected to be returned for evaluation.It has not yet been received.A follow up report will be submitted with all additional relevant information.
 
Event or Problem Description
It was reported that the device was not prepped (air aspiration) outside the anatomy prior to use.The procedure was to treat a lesion located in the circumflex artery that was mildly calcified.The nc trek rx balloon dilatation catheter was advanced to the lesion and the balloon ruptured during the first inflation at 8 atmospheres.There was no reported adverse patient effect and no clinically significant delay in the procedure.A non-abbott balloon was used to complete the procedure.No additional information was provided.
 
Additional Manufacturer Narrative
The device was returned for analysis.The reported material rupture was able to be confirmed.Additionally, scratches on the balloon and peeling balloon material were noted.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other incidents from this lot.It should be noted that the nc trek rx coronary dilatation catheter instructions for use (ifu), states: caution: all air must be removed from the balloon and displaced with contrast prior to inserting into the body otherwise, complications may occur.The investigation determined the reported difficulties appear to be related to circumstances of the procedure as it is likely that interaction with the mildly calcified anatomy resulted in the noted scratched and peeling balloon material; thus compromising the balloon material and ultimately resulting in the reported material rupture.There is no indication of a product quality issue with respect to manufacture, design or labeling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Common Device Name
CORONARY DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
MDR Report Key10744533
Report Number2024168-2020-08957
Device Sequence Number3006654
Product Code LOX
UDI-Device Identifier08717648151880
UDI-Public08717648151880
Combination Product (Y/N)N
Initial Reporter CountryCO
PMA/510(K) Number
K103153
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 12/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2022
Device Model Number1012449-20
Device Catalogue Number1012449-20
Device Lot Number90820G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Initial Date Received by Manufacturer 10/05/2020
Supplement Date Received by Manufacturer11/19/2020
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date12/07/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-