CATH TEMPO 5F PIG 100CM 5SH; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number 451503H5
Medical Device Problem Code Device Contamination with Chemical or Other Material (2944)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 09/24/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device history record (dhr) review was conducted and the product met quality requirements for product acceptance.This device is available for analysis but has not yet been received.Additional information is pending and will be submitted within 30 days upon receipt.
 
Event or Problem Description
During prep, a 5f 100cm pigtail (pig) 5 side holes (sh) tempo catheter tip was contaminated.There were difficulties removing the product from the packaging.The packaging did not remove.There was no reported patient injury.Another tempo catheter was used to complete the procedure.The target lesion is the carotid artery with no tortuosity, calcification, and twenty percent stenosis.The device was handled according to the instruction for use (ifu).There were no damages to the device packaging.The device will be returned for evaluation.No other information was provided.
 
Additional Manufacturer Narrative
During prep, a 5f 100cm pigtail (pig) 5 side holes (sh) tempo catheter tip was contaminated.There were difficulties removing the product from the packaging.The packaging did not remove.Another tempo catheter was used to complete the procedure.The target lesion is the carotid artery with no tortuosity, calcification, and twenty percent stenosis.The device was handled according to the instruction for use (ifu).There were no damages to the device packaging.No other information was provided.There was no reported patient injury.The device was returned for analysis.One non-sterile unit of a tempo diagnostic catheter (cath tempo 5f pig 100cm 5sh) was received for analysis.The original packaging was not returned.Per visual analysis, three kinks were observed on the catheter body/shaft approximately at 15.8 cm, 45.3 cm and 72.9 cm from the strain relief.No foreign material was observed on the unit.No other anomalies were observed on the unit.Per functional analysis, the flushing test was performed successfully.Neither an obstruction nor a foreign material was observed coming out of the device.Per dimensional analysis, the outer diameter and inner diameter of the body/shaft of the catheter were measured near every kink and the results were found within specification.A product history record (phr) review of lot 17770447 revealed no anomalies or non-conformances during the manufacturing and inspection processes that can be associated with the reported event.The reported ¿packaging/pouch/box - removal difficulty¿ was not confirmed since the original packaging was not returned with the unit.The reported ¿catheter (body/shaft) - foreign material - during prep¿ was not confirmed since no foreign material was observed on the unit.Flushing test was performed on the unit and no foreign material was observed coming out of the unit.Dimensional test near every observed kink was performed and results were observed within specification.However, the exact cause of the kinked areas on the unit could not be conclusively determined during the analysis.Storage/handling factors may have led to the reported event.Per the instructions for use (ifu), which is not intended as a mitigation of risk, ¿do not expose to organic solvents.Store in a cool, dark, dry place.Do not use if package is open or damaged.Do not use the catheter if the ¿use by¿ date on the package label has expired.Do not resterilize.Exposure to temperatures above 54oc (130of) may damage the catheter.¿ neither the phr review, product analysis, nor the information available suggests a design or manufacturing related cause for the reported events therefore, no corrective/preventive action will be taken at this time.
 
Additional Manufacturer Narrative
This device was received for analysis, but the engineering report is not yet available.However, it will be submitted within 30 days upon receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CATH TEMPO 5F PIG 100CM 5SH
Common Device Name
CATHETER, INTRAVASCULAR, DIAGNOSTIC
MDR Report Key10744626
Report Number9616099-2020-04019
Device Sequence Number11541153
Product Code DQO
UDI-Device Identifier10705032008273
UDI-Public10705032008273
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K973401
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2021
Device Model Number451503H5
Device Catalogue Number451503H5
Device Lot Number17770447
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/10/2020
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer11/10/2020
12/03/2020
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date12/07/2020
12/09/2020
Was Device Evaluated by Manufacturer? (Y/N) Device Not Returned to Manufacturer
Date Device Manufactured04/06/2018
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Weight56
-
-