Model Number 420230-04
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Isi has not yet received the large hem-o-lok clip applier instrument for evaluation.Therefore, the root cause of the customer reported failure mode has not been determined.A follow-up mdr will be submitted if the product is returned and evaluated and/or if additional information is received.A log review was performed for the large hem-o-lok clip applier instrument reported in this complaint (pn: 420230-04/m10110606661) and the following was found: the instrument was last used on 01-jul-2020 on system sh1249.The instrument had 6 uses remaining and was not used for any subsequent procedures.No error would appear in the logs for this type of reported issue.No photo, or video was provided by the site for review.This complaint is being reported because a large hem-o-lok clip applier instrument failed to fully latch a clip closed or incurred a failure mode that is known to impact clip application effectiveness.Deficiencies in clip application may lead to inadequate hemostasis.While there was no harm, or injury to the patient, the reported failure mode could likely cause, or contribute to an adverse event if it were to recur.The product is not implantable.It is unknown if the initial reporter submitted a report to the fda.
 
Event or Problem Description
It was reported that during a da vinci-assisted surgical procedure, the customer stated that the large hem-o-lok clip applier instrument was unable to work and it cannot hold the clip correctly.The procedure was completed with no reported injury.No further details have been received as of the date of this report.
 
Additional Manufacturer Narrative
Additional information can be found in the following sections: d10, g4, g7, h2,h3, h6, and h10.4307 intuitive surgical, inc.(isi) received the large hem-o-lok clip applier instrument involved with this complaint and completed the device evaluation.Failure analysis (fa) investigation confirmed the customer reported complaint.The instrument was found to have a loose grip cable at the distal end.The root cause of this failure is attributed to a component failure.In addition, fa found that the instrument had a broken grip cable at the proximal end.The root cause of this failure is attributed to a component failure.
 
Event or Problem Description
Refer to h10/h11 for follow-up information.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDOWRIST;DAVINCI SI
Common Device Name
LARGE HEM-O-LOK CLIP APPLIER
Manufacturer (Section D)
INTUITIVE SURGICAL, INC
3410 central expressway
santa clara CA 95051
MDR Report Key10744697
Report Number2955842-2020-11130
Device Sequence Number19630402
Product Code NAY
UDI-Device Identifier00886874111680
UDI-Public(01)00886874111680(10)M10110606
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K050369
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,other
Initial Reporter Occupation Non-Healthcare Professional
Remedial Action Other
Type of Report Initial,Followup
Report Date (Section B) 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number420230-04
Device Catalogue Number420230
Device Lot NumberM10110606 661
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/13/2020
Initial Date Received by Manufacturer 10/01/2020
Supplement Date Received by Manufacturer11/16/2020
Initial Report FDA Received Date10/27/2020
Supplement Report FDA Received Date12/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
DA VINCI INSTRUMENTS AND ACCESSORIES.
-
-