Model Number 71364451
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 09/30/2020
Event Type  malfunction  
Event Description
It was reported that the depth gauge was found broken after washing cycle.No case involved.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The associated device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The device was found to be fractured.The fractured pieces were not returned.The device was manufactured in 2013 and shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
R3 DEPTH GAUGE
Type of Device
APPLIANCE,FXTIN,NAIL/BLD/PLATE CMBNTIN,MULTIPLE COMPONENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10744993
MDR Text Key213379616
Report Number1020279-2020-05858
Device Sequence Number1
Product Code KTT
UDI-Device Identifier03596010599438
UDI-Public03596010599438
Combination Product (y/n)N
PMA/PMN Number
K093047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received10/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364451
Device Catalogue Number71364451
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received01/08/2021
Is This a Reprocessed and Reused Single-Use Device? No