DEPUY ORTHOPAEDICS INC US LPS UNIV TIB HIN INS SM 12MM; LPS AND S-ROM : KNEE TIBIAL INSERT Back to Search Results
Model Number 1987-27-212
Medical Device Problem Code Detachment of Device or Device Component (2907)
Health Effect - Clinical Codes Inflammation (1932); Pain (1994); No Code Available (3191)
Date of Event 10/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Change of lps small insert.Even though the insert seems not broken, there is a internal "shift between poly and" hinge part.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device was not able to confirm the reported event.The investigation found no evidence of product malfunction or product error and the need for corrective action was not established.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : a worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the provided lot code bfa0n0n is unrecognized in the unity master data, as400 and jde.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d10; corrected: h3.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LPS UNIV TIB HIN INS SM 12MM
Common Device Name
LPS AND S-ROM : KNEE TIBIAL INSERT
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10747593
Report Number1818910-2020-23430
Device Sequence Number19670211
Product Code KRO
UDI-Device Identifier10603295079484
UDI-Public10603295079484
Combination Product (Y/N)N
Initial Reporter CountryZA
PMA/510(K) Number
K091453
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1987-27-212
Device Catalogue Number198727212
Device Lot NumberBFAONON
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/16/2020
Initial Date Received by Manufacturer 10/13/2020
Supplement Date Received by Manufacturer11/16/2020
12/15/2020
Initial Report FDA Received Date10/28/2020
Supplement Report FDA Received Date11/19/2020
12/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age29 YR
-
-