| Catalog Number |
0210100000 |
| Medical Device Problem Code |
Delivered as Unsterile Product (1421)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
09/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device was received.Event confirmation status: 1 reported event was confirmed.Evaluation results: 1 device did not have a root cause established.Additional information: 1 device was labeled for single-use.1 device was not reprocessed or reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 1 malfunction event in which the device had debris in sterile package.1 event had patient involvement; no patient impact.
|
| |
|
Search Alerts/Recalls
|