Model Number 393014
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 10/21/2020
Type of Reportable Event Serious Injury
Event or Problem Description
Device was explanted and replaced with a smaller device because the patient was concerned that the device was too big and might come out of the pocket.Should additional information become available, this report will be updated.
 
Additional Manufacturer Narrative
The device was received and analyzed.The memory content of the icd was inspected, showing a normal device function while implanted and in service.The icd was subjected to an electrical analysis.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a correct sensing and shock delivery.In particular, the specified energy level was reached.In conclusion, the device proved to be fully functional.The analysis of the memory content showed a normal device behavior while the device was implanted and in service.There was no indication of a device malfunction.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ITREVIA 7 HF-T DF-1
Common Device Name
CRT-D
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
MDR Report Key10748873
Report Number1028232-2020-04614
Device Sequence Number19777459
Product Code NIK
UDI-Device Identifier04035479129309
UDI-Public04035479129309
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P050023
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date09/30/2018
Device Model Number393014
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/03/2020
Initial Date Received by Manufacturer 10/22/2020
Supplement Date Received by Manufacturer09/10/2021
Initial Report FDA Received Date10/28/2020
Supplement Report FDA Received Date09/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age87 YR
-
-