| Model Number |
393014 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
10/21/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
Device was explanted and replaced with a smaller device because the patient was concerned that the device was too big and might come out of the pocket.Should additional information become available, this report will be updated.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was received and analyzed.The memory content of the icd was inspected, showing a normal device function while implanted and in service.The icd was subjected to an electrical analysis.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed.A fibrillation signal was applied and the device delivered a defibrillation shock as specified, documenting a correct sensing and shock delivery.In particular, the specified energy level was reached.In conclusion, the device proved to be fully functional.The analysis of the memory content showed a normal device behavior while the device was implanted and in service.There was no indication of a device malfunction.
|
| |
|