• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. RENASYS EZ MAX HOSPITAL; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. RENASYS EZ MAX HOSPITAL; NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP Back to Search Results
Model Number 66801309
Medical Device Problem Code Defective Alarm (1014)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 09/29/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during service evaluation the device was not able to generate alarms under expected alarm conditions.No case involved.
 
Additional Manufacturer Narrative
The device intended for use in treatment was returned for evaluation and we have established a relationship between the device and the reported event.A visual inspection found no defects.The functional evaluation found that the device failed to charge; could not generate alarms under expected alarm conditions and the root cause identified as a defective mainboard.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history found other related events.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew are taking further actions relating to the failure reported and continue to monitor for adverse trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RENASYS EZ MAX HOSPITAL
Common Device Name
NEGATIVE PRESSURE WOUND THERAPY POWERED SUCTION PUMP
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull HU3 2 BN
UK  HU3 2BN
MDR Report Key10749533
Report Number8043484-2020-03298
Device Sequence Number14190420
Product Code OMP
UDI-Device Identifier00040565126470
UDI-Public00040565126470
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K132446
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 04/09/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number66801309
Device Catalogue Number66801309
Initial Date Received by Manufacturer 09/29/2020
Supplement Date Received by Manufacturer03/30/2021
Initial Report FDA Received Date10/28/2020
Supplement Report FDA Received Date04/09/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-