| Model Number |
66801309 |
| Medical Device Problem Code |
Defective Alarm (1014)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
09/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that during service evaluation the device was not able to generate alarms under expected alarm conditions.No case involved.
|
| |
|
Additional Manufacturer Narrative
|
|
The device intended for use in treatment was returned for evaluation and we have established a relationship between the device and the reported event.A visual inspection found no defects.The functional evaluation found that the device failed to charge; could not generate alarms under expected alarm conditions and the root cause identified as a defective mainboard.The manufacturing records show no evidence that the product did not meet specification at the time of manufacture.The complaint history found other related events.This investigation is now complete with no further action deemed necessary at this stage.Smith + nephew are taking further actions relating to the failure reported and continue to monitor for adverse trends.
|
| |
|