|
It was reported that the drill bit was used in a total hip replacement procedure to drill a hole for reflection screw, after three uses the drill bent.Procedure finished with a smith and nephew backup device.Patient injuries were not reported.Surgical delay less than 30 minutes.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
The associated device, intended to use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.A relationship, if any, between the device and the reported incident could not be corroborated.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
|