• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US MICRO TORNADO HP W HANDCONTROL; LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM Back to Search Results
Model Number 283512
Device Problems Break (1069); Electrical Shorting (2926)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 09/14/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint #: (b)(4).Investigation summary: the device was received and evaluated at the service center.The reported complaint that the hand piece was not working was confirmed.On evaluation, it was further found that the thread of the hand piece was broken and that the hand piece was not recognized by the fms system.It was also found that there was a short circuit in the motor and that the tissue seal was missing.The device was modified as per the guidelines of the manufacturer, repaired, tested and found to be fully functional.User mishandling of the device is the most probable root cause of the broken thread and the missing tissue seal.Fluid ingress into the motor is the most probable root cause of the short circuit in the motor, which in turn, would have caused the hand piece to not function as reported by the customer.The service history has been reviewed in lieu of the device history record for this device since it was previously serviced.The device was last serviced on 05/06/2019 and passed all functional testing before being returned to the customer.At this point in time, no corrective action is required, and no further action is warranted.Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.Udi: (b)(4).
 
Event Description
It was reported by the affiliate via complaint submission tool, two handpieces of fms systems were not working.Additional information could not be provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO TORNADO HP W HANDCONTROL
Type of Device
LINE-POWERED ARTHROSCOPIC SHAVER SYSTEM
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
Manufacturer (Section G)
DEPUY MITEK
325 paramount drive
raynham MA 02767
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key10749585
MDR Text Key213697372
Report Number1221934-2020-03211
Device Sequence Number1
Product Code HRX
UDI-Device Identifier10886705016938
UDI-Public10886705016938
Combination Product (y/n)N
Reporter Country CodeLH
PMA/PMN Number
K954465
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 10/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283512
Device Catalogue Number283512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/25/2020
Initial Date Manufacturer Received 10/13/2020
Initial Date FDA Received10/28/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Type of Device Usage N
Patient Sequence Number1
-
-