• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED SCREW 105MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TFNA FENESTRATED SCREW 105MM - STERILE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 04.038.205S
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Injury (2348)
Date of Event 10/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, while the surgeon was inserting proximal femoral nailing system (tfna) fenestrated lag screw it was determine that she had perforated the head of the femur as per x-rays result.The tfna fenestrated lag screw was removed and repositioned to a more favorable place in the femoral head and neck and surgery.It was unknown if there was a surgical delay.The procedure was successfully completed.Patient outcome was unknown.Concomitant device reported: tfna nail (part# 04.037.042s, lot# unknown, quantity 1).This complaint involves one (1) device.This report is for one (1) tfna fenestrated screw 105mm - sterile.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TFNA FENESTRATED SCREW 105MM - STERILE
Common Device Name
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10750581
Report Number2939274-2020-04856
Device Sequence Number3058940
Product Code HSB
UDI-Device Identifier10886982098795
UDI-Public(01)10886982098795
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.038.205S
Device Catalogue Number04.038.205S
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 10/01/2020
Initial Report FDA Received Date10/28/2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-