• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.75/15; CORONARY DRUG-ELUTING STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK AG, BUELACH, SWITZERLAND ORSIRO (US) 2.75/15; CORONARY DRUG-ELUTING STENT Back to Search Results
Model Number 401743
Medical Device Problem Code Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 10/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Combination product: yes.
 
Event or Problem Description
An orsiro drug-eluting stent system was selected for treatment.The device could not be advanced.Another orsiro was implanted.
 
Additional Manufacturer Narrative
Combination product: yes.Neither the complaint instrument nor the angiographic material was returned for analysis.Therefore, no technical investigation on the subject could be performed.The production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.Review of the production documentation for the product detailed above verified that the instrument was manufactured according to specifications and fulfilled all the requirements of in-process and final inspection.During final inspection, every stent system undergoes visual inspection to ensure correct stent embedding and homogeneous crimping of the stent.Further, the stent outer diameter is verified to 100 percent.Based on the conducted investigations of the device being subject to this complaint, no material or manufacturing related root cause was determined.It should be noted that, according to the ifu, pre-dilatation of the lesion is mandatory.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ORSIRO (US) 2.75/15
Common Device Name
CORONARY DRUG-ELUTING STENT
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10750673
Report Number1028232-2020-04690
Device Sequence Number3777324
Product Code NIQ
UDI-Device Identifier07640130438989
UDI-Public07640130438989
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P170030
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/16/2021
Device Model Number401743
Device Catalogue NumberSEE MODEL NO.
Device Lot Number05196221
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 10/09/2020
Supplement Date Received by Manufacturer12/08/2020
Initial Report FDA Received Date10/28/2020
Supplement Report FDA Received Date12/08/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age86 YR
-
-